Skip to the content.

Your Data

A download from the configuration dashboard gives you two files per device:

File What it is
<name>_<start>.ttg The raw log, byte for byte as the device sent it. This is your archival copy.
<name>.pkl The same data parsed into Python objects, ready for analysis.

<name> is the device’s label, or its ID if it has none, with any characters a file name cannot hold left out. <start> is when the study started, so downloading from a device twice simply replaces the file. Both tools name a device’s .ttg the same way.

The browser tool saves only the .ttg. A .pkl is a Python file, made by the desktop dashboard.

Keep the .ttg files. They are the original record and can be re-parsed at any time; the .pkl is a convenience that can be rebuilt from its .ttg whenever you need it, including from a .ttg the browser tool saved:

tottag --convert ~/Downloads/02_1788300000.ttg

This writes the same 02.pkl the download wrote, next to the .ttg file. Keep the name the .ttg file was saved under, because that is how the tool knows which device’s log it is. If a device’s data ever looks wrong, the .ttg file is what tells you whether the problem is in the recording or in the processing.


Reading a log in Python

import pickle
records = pickle.load(open('02.pkl', 'rb'))

You get a list of dictionaries, one per instant, sorted by time. Every record has a t (a Unix timestamp) plus whichever fields were recorded at that moment:

Key Contents
r Distances. {device_id: millimetres} — the main data
v Battery voltage, mV
c Charging event — 'Plugged', 'Unplugged', 'Charging', 'Not Charging'
m Motion — True when moving
i Accelerometer, [x, y, z]
b Other devices seen over Bluetooth
rst A restart, with its cause as a list of strings
rtc, offset, lag Timekeeping — see clock drift
diag Internal health counters, which firmware build wrote the log, the chip temperature, and radio health

So to retrieve the distances over time to device 0x3E:

series = [(r['t'], r['r'][0x3E]) for r in records if 'r' in r and 0x3E in r['r']]

The dashboard’s processing module (from tottag.processing import *) has helpers for plotting pairs and overlaying annotated events; see The Dashboard.


How precise is a distance?

Between two devices with a clear path, precision is excellent. Across a ten-device deployment, every pair agreed with itself from both directions to within 1 mm of the median. In practice:

The spread is the useful signal. A pair with a stable distance has a clear path; a pair whose distance jumps around has something between them. Compute the standard deviation over a short window and treat high-variance stretches as “obstructed” rather than as noise to be smoothed away.

One caveat

A missing reading is not a distance of zero. If two devices could not complete an exchange, there is simply no entry for that pair at that instant. Absence usually means far apart or obstructed, but it can also mean one device was charging, asleep, or had left the network. Check for charging events and restarts before reading too much into a gap.


Checking that a log is complete

Worth doing before you reuse the devices.

1. The download reported no warnings. A clean download prints nothing unusual. Lines about retransmission rounds are normal and self-correcting. A WARNING: log from ... is incomplete is not — download from that device again.

2. The dates cover your study.

import datetime
ts = [r['t'] for r in records]
print(datetime.datetime.fromtimestamp(ts[0]), '->', datetime.datetime.fromtimestamp(ts[-1]))

3. Every device saw every other device.

import collections
peers = collections.Counter()
for r in records:
    if 'r' in r:
        peers.update(r['r'].keys())
print({hex(k): v for k, v in peers.items()})

In a ten-device study each log should list nine peers. A device missing entirely was never in the network. A device with far fewer readings than the others was on the edge of range — real data, but treat it as a weaker measurement.

4. Restarts are explained.

for r in records:
    if 'rst' in r:
        print(datetime.datetime.fromtimestamp(r['t']), r['rst'])

SW Power-On entries are normal — a device restarts on every plug and unplug, so these should line up with your charging routine. Anything mentioning Watchdog is a fault; see Troubleshooting.


Gaps in the data

Not every gap is a problem, and the kind of gap tells you which.

Gap Meaning
Distances stop, housekeeping continues every 5 min The device was running but had nobody in range. Normal when people separate.
Everything stops, then a Plugged event On a charger or a USB cable. Recording is suspended while plugged in.
Everything stops, then a restart Powered off — outside the study window, or dead battery. Check the voltage before the gap.
Everything stops with no explanation Investigate. Start with Troubleshooting.

The 5-minute housekeeping cadence is what makes the first two distinguishable, so it is worth checking which you have before concluding a device failed.


Daily figures

Both tools can sum a deployment into daily figures for chosen devices: the browser tool on its Time together tab, and get_daily_ranging_statistics in the desktop dashboard’s processing module.

The browser tool’s CSV export has the columns date, tag, with, minutes_in_range, minutes_within_<n>_<unit>, mean_distance_<unit>, rounds_in_range, rounds_within_touching, cutoff_<unit>, touching_<unit>, and time_zone.


Clock drift

Devices keep time with a crystal that is not perfect. Measured across deployments, they run roughly 160–190 parts per million slow — about 15 seconds per day.

Every device in a deployment follows the same network clock, so devices agree with each other to within a few parts per million. The drift is almost entirely common to the whole group.

This matters only if you are aligning TotTag data against an external source — video, an observer’s notes, another instrument. Over a week, expect roughly a minute and a half of offset against wall clock. The rtc, offset, and lag fields in each 5-minute housekeeping record carry the device’s own clock alongside the network clock, so the drift can be measured and corrected after the fact rather than estimated.


The raw format

The .ttg file is a page-framed stream: a header, then a sequence of pages each carrying its own sequence number, timestamps, and CRC. The CRC is what lets the tools detect a page damaged in transfer and ask the device to resend it.

You rarely need this level of detail. If you are writing your own reader, the authoritative description is the two existing readers in the tottag-dashboard repository — Python/dashboard/tottag_format.py and Web/packages/tottag-schema/src/log.ts — which are kept in step with the firmware by an automated check.